DCV |
Maximum Voltage: 1000V Accuracy: Fluke 83 V: ±(0.1%+1) / Fluke 87 V: ±(0.05%+1) Maximum Resolution: Fluke 83 V: 100 µV / Fluke 87 V: 10 µV |
|
ACV |
Maximum Voltage: 1000V Accuracy: Fluke 83 V: ±(0.5%+2) / Fluke 87 V: ±(0.7%+2) True RMS AC Bandwidth: Fluke 83 V: 5kHz Fluke 87 V: 20kHz* / *with low pass filter; 3db @ 1kHz Maximum Resolution: 0.1 mV |
|
DC Current |
Maximum Amps: 10A Amps Accuracy: Fluke 83 V: ±(0.4%+2) / Fluke 87 V: ±(0.2%+2) Maximum Resolution: Fluke 83 V: 0.01 mA / Fluke 87 V: 0.01 µA |
|
AC Current |
Maximum Amps: 10A Amps Accuracy: Fluke 83 V: ±(1.2%+2) / Fluke 87 V: ±(1.0%+2) True RMS Maximum Resolution: 0.1 µA |
|
Resistance |
Maximum Resistance: 50 MW Accuracy: Fluke 83 V: ±(0.4%+1) / Fluke 87 V: ±(0.2%+1) Maximum Resolution: 0.1W |
|
Capacitance |
Maximum Capacitance: 9,999 µF Accuracy: ±(1%+2) Maximum Resolution: 0.01 nF |
|
Frequency |
Maximum Frequency: 200 KHz Accuracy: ±(0.005%+1) Maximum Resolution: 0.01 Hz |
|
Duty Cycle |
Maximum Duty Cycle: 99.9% Accuracy: ±(0.2 per KHz +0.%) Maximum Resolution: 0.1% |
|
Temperature Measurement |
87 V, 87 V/E
-200.0°C - 1090°C
-328.0°F - 1994.0°F, 1.0% excluding probe |
|
80 BK Temperature Probe |
Probe -40 °C - 260 °C
-40 °F - 500 °F, 2.2 °C or 2% whichever is greater |
|
Conductance |
Maximum Conductance: 60.00nS Accuracy: ±(1.0%+10) Maximum Resolution: 0.01nS |
|
Operating Range |
Operating Temperature: -20 °C - 55 °C
Storage Temperature: -40 °C - 60 °C |
|
Physical |
5.2 cm x 9.8cm x 20.1 cm (2.06 x 3.86 x 7.93) / 624 g (22 oz.) |
|
Included Accessories |
87V and 83V; TL75 test leads, AC72 Alligator Clips, Holster, 9V battery (installed), Getting Started Manual, and CD with detailed manual. 87V includes an 80BK temperature probe.
87V/E; 87V, TL224 silicone test leads, AC220 long reach alligator clips, TP238 test probes, magnet and strap, 80BK temperature probe, holster, 9V battery (installed), Getting Started Manual, and CD with detailed manual.
|