| FLUKE - 725Ex CALIBRAÇÃO DE PROCESSO |
|
Voltage DC |
30.000 V - 0.02%+ 2 counts (upper display)
30.000 V - 0.02%+ 2 counts (lower display)
100.00 mV - 0.02%+ 2 counts
-10.00 mV to 75.00 mV - 0.025 % + 1 count (via TC connector) |
|
Current DC |
24.000 mA - 0.02%+ 2 counts |
|
Resistance |
0.0 to 400.0 OHM 0.1 OHM (4-wire), 0.15O (2- and 3-wire)
401 to 1500 OHM 0.5 OHM (4-wire), 1O (2- and 3-wire)
1500 to 3200 OHM 1 OHM (4-wire), 1.5O (2- and 3-wire |
|
Frequency |
2,0 a 1000,0 CPM - 0,05% + 1 count
1,0 a 1100,0 Hz - 0,05% + 1 count
1,00 a 10,00 kHz - 0,05% + 1 count
Sensitivity - 1 V peak-to-peak-minimum |
|
Pressure |
Accuracy - from 0.025% of range using any of 8 intrinsic safe pressure modules.
(for detailed specifications refer to pressure modules in options and accessories)
Modules available for differential, gage, vacuum, absolute, dual and high pressure. |
|
Voltage DC |
100,00 mV - 0,02% + 2 counts
10,000 V - 0,02% + 2 counts
-10,00 mV to 75,00 mV - 0,025% + 1 count (via TC connector) |
|
Current DC |
24,000 mA (Source) - 0,02% + 2 counts
24,000 mA (Simulate) - 0,02% + 2 counts |
|
Resistance |
15,0 to 400,0 OHM - 0,15 OHM (exc. current 0,15 a 0,5 mA), 0,1 OHM (exc. current 0,5 a 2 mA)
401 to 1500 OHM - 0,5 OHM (excitation current 0,05 a 0,8 mA)
1500 to 3200 OHM - 1 OHM (excitation current 0,05 a 0,4 mA) |
|
Frequency |
2,0 to 1000,0 CPM - 0.05%
1,0 to 1100,0 Hz - 0.05%
1,00 to 10,00 kHz - 0.25%
Waveform: 5 V p-p squarewave, -0.1 V offset |
|
Measure Accuracy: |
NI-120 - 0,2 °C
PT-100 - (385) 0,33 °C
PT-100 - (393) 0,3 °C
PT-100 - (JIS) 0,3 °C
PT-200 - (385) 0,2 °C
PT-500 - (385) 0,3 °C
PT-1000 - (385) 0,2 °C
Resolution: 0,1 °C
J: 0,7 °C / K: 0,8 °C / T: 0,8 °C / E: 0,7 °C / R: 1,8 °C / S: 1,5 °C / B: 1,4 °C / L: 0,7 °C / U: 0,75 °C / N: 0,9 °C
Resolution J, K, T, E, L, N, U: 0,1 °C, 0,1 °F B, R, S: 1 °C, 1 °F |
|
Source Accuracy: |
NI-120 - 0,2 °C
PT-100 (385) - 0,33 °C
PT-100 (393) - 0,3 °C
PT-100 (JIS) - 0,3 °C
PT-200 (385) - 0,2 °C
PT-500 (385) - 0,3 °C
PT-1000 (385) - 0,2 °C
Resolution:0,1 °C
Note: Accuracy stated for 4-wire measurement.
J: 0,7 °C / K: 0,8 °C / T: 0,8 °C / E: 0,7 °C / R: 1,4 °C / S: 1,5 °C / B: 1,4 °C / L: 0,7 °C / U: 0,75 °C / N: 0,9 °C
Resolution J, K, T, E, L, N, U: 0,1 °C, B, R, S: 1 °C |
|
Ramp Functions |
Source functions: Voltage, current, resistance, frequency, temperature
Ramps - Slow ramp, Fast ramp, 25% step-ramp |
| Loop Power Function |
Voltage: 12 V
Accuracy: 10%
Maximum current: 22 mA, short circuit protected |
| Step Functions |
Source functions: Voltage, current, resistance, frequency, temperature
Steps - 25% of range, 100% of range |
| Agency Approvals |
ATEX II 1 G EEx ia IIB 171°C
I.S. Class I, Division 1 Groups |
|
MECHANICAL & GENERAL SPECIFICATIONS |
|
Size |
130 x 236 x 61 mm
5.188" x 9.291" x 2.402" |
| Weight |
0.85 kg (1.874 lbs.) |
| Batteries |
4 AA alkaline batteries |
| Battery Replacement |
Separate battery compartment, accessible without breaking calibration seal |
| Side port connections |
Pressure module connector |
|
 |
| MULTITEC - Instrumentos de Medição Ltda. |
Rua Pedro Ernesto, 70-Lj - Saúde - Rio de Janeiro - RJ CEP:. 20220-350
TEL: (0XX21) 2233-3283 / 2516-2153 :: FAX: (0XX21) 2233-3847 |
|
 |
|
|
|
|